The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Metrics
Affected Vendors & Products
References
History
Mon, 18 Nov 2024 16:15:00 +0000
Type | Values Removed | Values Added |
---|---|---|
First Time appeared |
Baxter
Baxter life2000 Ventilator Firmware |
|
CPEs | cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:* | |
Vendors & Products |
Baxter
Baxter life2000 Ventilator Firmware |
|
Metrics |
ssvc
|
Thu, 14 Nov 2024 21:45:00 +0000
Type | Values Removed | Values Added |
---|---|---|
Description | The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. | |
Title | Debug port on Life2000 Ventilator serial interface is enabled by default | |
Weaknesses | CWE-1263 | |
References |
| |
Metrics |
cvssV3_1
|

Status: PUBLISHED
Assigner: Baxter
Published:
Updated: 2024-11-18T15:33:44.924Z
Reserved: 2024-10-10T19:24:41.495Z
Link: CVE-2024-48973

Updated: 2024-11-18T15:33:39.812Z

Status : Awaiting Analysis
Published: 2024-11-14T22:15:18.510
Modified: 2024-11-15T13:58:08.913
Link: CVE-2024-48973

No data.