A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Improper Access Control could allow an unauthenticated attacker to escape from the restricted environment (“kiosk mode”) and access the underlying operating system. Successful exploitation requires direct physical access to the system.
Metrics
Affected Vendors & Products
References
History
No history.

Status: PUBLISHED
Assigner: siemens
Published:
Updated: 2024-08-04T13:30:21.872Z
Reserved: 2020-07-15T00:00:00
Link: CVE-2020-15797

No data.

Status : Modified
Published: 2020-10-13T16:15:21.280
Modified: 2024-11-21T05:06:12.010
Link: CVE-2020-15797

No data.