Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
Metrics
Affected Vendors & Products
References
Link | Providers |
---|---|
https://www.us-cert.gov/ics/advisories/icsma-20-170-04 |
![]() ![]() |
History
No history.

Status: PUBLISHED
Assigner: icscert
Published:
Updated: 2024-08-04T11:48:58.109Z
Reserved: 2020-04-21T00:00:00
Link: CVE-2020-12039

No data.

Status : Modified
Published: 2020-06-29T14:15:11.617
Modified: 2024-11-21T04:59:09.853
Link: CVE-2020-12039

No data.