The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Metrics
Affected Vendors & Products
References
History
No history.

Status: PUBLISHED
Assigner: icscert
Published:
Updated: 2024-08-06T11:41:49.223Z
Reserved: 2014-08-22T00:00:00
Link: CVE-2014-5406

No data.

Status : Modified
Published: 2015-07-06T19:59:00.097
Modified: 2024-11-21T02:11:59.410
Link: CVE-2014-5406

No data.