Search Results (22 CVEs found)

CVE Vendors Products Updated CVSS v3.1
CVE-2021-4481 1 Draeger 1 Protector Software 2026-06-03 8.2 High
Dräger Protector Software prior to version 6.4.2 contains a local privilege escalation vulnerability due to insecure file system permissions that allows local attackers to execute arbitrary code with elevated privileges. Attackers can replace binaries or loaded modules on the host system to execute code with NT SYSTEM privileges.
CVE-2019-25718 1 Draeger 1 Infinity Explorer C700 2026-06-03 8.4 High
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
CVE-2019-25716 1 Draeger 2 Infinity Delta, Infinity Kappa 2026-06-03 6.5 Medium
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
CVE-2025-15653 1 Draeger 2 Zeus Ie, Zeus Rs C500 2026-06-03 6.8 Medium
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
CVE-2024-14036 1 Draeger 2 Core, M540 Converter Service 2026-06-03 7.5 High
Dräger Core 1.0.5 and Dräger M540 Converter Service 1.0.9 contain a denial of service vulnerability that allows network-adjacent attackers to trigger high CPU load by sending specially crafted, unencrypted SDC messages during the discovery process. Attackers with access to the hospital network can send malformed SDC packets to exhaust CPU resources in the affected process, causing further SDC messages to no longer be processed.
CVE-2022-4992 1 Draeger 2 Infinity Acute Care System, Standalone Infinity M540 Patient Monitor 2026-06-03 8.6 High
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.
CVE-2021-4480 1 Draeger 1 Protector Software 2026-06-03 8.2 High
Dräger Protector Software prior to version 6.4.2 contains a local privilege escalation vulnerability due to insecure file system permissions that allows local attackers to execute arbitrary code with elevated privileges. Attackers can replace binaries or loaded modules on the host system to execute code with NT SYSTEM privileges.
CVE-2021-4479 1 Draeger 1 Atlan A350 2026-06-03 4 Medium
Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
CVE-2021-4478 1 Draeger 2 Cc-vision Basic, Cc-vision E-cal 2026-06-03 8.2 High
Dräger CC-Vision Basic before 7.5.3 and Dräger CC-Vision E-Cal before 7.2.5.0 contain an out-of-bounds write vulnerability when loading .gdt files. A crafted .gdt file can trigger a buffer overflow during file parsing, allowing an attacker to crash the application or execute malicious code on the underlying system.
CVE-2019-25724 1 Draeger 1 Infinity M300 2026-06-03 6.5 Medium
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
CVE-2019-25723 1 Draeger 1 Perseus A500 2026-06-03 4 Medium
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
CVE-2019-25722 1 Draeger 5 Sc6802xl, Sc8000, Sc90000 Xl and 2 more 2026-06-03 7.6 High
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
CVE-2019-25721 1 Draeger 1 Infinity M300 2026-06-03 6.5 Medium
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
CVE-2019-25719 1 Draeger 2 Infinity Acute Care System, Standalone Infinity M540 Patient Monitor 2026-06-03 8.6 High
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
CVE-2019-25717 1 Draeger 2 Infinity Delta, Infinity Kappa 2026-06-03 4.3 Medium
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.
CVE-2025-2810 1 Draeger 1 Icmhelper 2026-04-15 5.5 Medium
A low privileged local attacker can abuse the affected service by using a hardcoded cryptographic key.
CVE-2025-41698 1 Draeger 1 Icmhelper 2026-04-15 7.8 High
A low privileged local attacker can interact with the affected service although user-interaction should not be allowed.
CVE-2021-28112 1 Draeger 4 X-dock 5300, X-dock 6300, X-dock 6600 and 1 more 2024-11-21 8.8 High
Draeger X-Dock Firmware before 03.00.13 has Active Debug Code on a debug port, leading to remote code execution by an authenticated attacker.
CVE-2021-28111 1 Draeger 4 X-dock 5300, X-dock 6300, X-dock 6600 and 1 more 2024-11-21 8.8 High
Draeger X-Dock Firmware before 03.00.13 has Hard-Coded Credentials, leading to remote code execution by an authenticated attacker.
CVE-2018-19014 1 Draeger 8 Delta Xl, Delta Xl Firmware, Infinity Delta and 5 more 2024-11-21 N/A
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.